The Vilex MCHI (Mini Cannulated Hemi Implant) and MMCI (Mini Metatarsal Cannulated Implant) are intended for the replacement of the base of the phalanx or the metatarsal head in lesser MPJs. The design is based on the successful performance of CHI for the big toe. The MCHI is made of titanium and the MMCI is made of cobalt chrome. The surface is spherical, concave or convex as appropriate. The main indications are a deteriorated joint causing pain, and lack of articulation.
*Patent Pending
Post-Operative Considerations
Patient should be warned that the prosthesis device is not a substitute for the natural MPJ and will
not enable the patient to function as if the joint is restored. The patient should avoid an active lifestyle. Excessive motion will accelerate the un- avoidable wear and tear and may loosen the prosthesis. Serious consequences such as fracture or shattering of the phalanx, metal fatigue, or dislocation of the device are possible.
Sterilization
The implant is provided non-sterile. Steam-sterilize the implant tray, implants and instruments, at 270°F for 10 minutes.
Available Sizes: MCHI and MMCI
| Product Number | MCHI-09/MMCI-09 | MCHI-10/MMCI-10 | MCHI-11/MMCI-11 | MCHI-12/MMCI-12 | MCHI-13/MMCI-13 |
|---|---|---|---|---|---|
| Diameter | 9 | 10.0 | 11.0 | 12.0 | 13.0 |
| Stem L | 11.5 | 11.5 | 12.0 | 12.0 | 12.0 |
| Thread D | 2.5 | 3.0 | 3.0 | 3.0 | 3.0 |
Before Proceeding
Without exception or equivocation, Vilex does not recommend the replacement of both the base of the phalanx and the metatarsal head in the same joint. The surgeon should evaluate the overall health of the patient and should consider the patient’s expectations before recommending implant surgery. The patient should be informed that a joint prostheses may loosen, cause fracture, or infection. Some patients exhibit sensitivity to cobalt chrome or titanium.
Implanting Procedure
• Make a longitudinal incision dorsally on MPJ
• Resect hypertrophic bone
• Free joint completely
• Resect & decompress joint
• Select trial sizer
• Drive a 1.1mm K-wire through sizer
• Remove sizer and slide implant over wire
• Engage driver into implant.
• Screw implant
• Stop advancing once implant is seated.

For more information about this and other Vilex products, contact: info@vilex.com.
Download Reorder Forms
FDA Authorization
Click on the link below to go to the United States Department of Health and
Human Services FDA 510(k) website.
Limited Warranty
Vilex warrants that this product meets the manufacturer's specifications and is free from manufacturing defects at the time of delivery. This warranty specifically excludes defects resulting from misuse, abuse or improper handling of the product subsequent to receipt by the purchaser. Vilex does not warrant the outcome of the surgical procedure.
Call us today for a free educational CD.